Text from the interactive explanation
Closed-loop high-precision two-photon lithography based on a multiplexed single-cavity dual-comb laser
Yu Y, Zhu Z, Willenberg B, Pupeikis J, Phillips CR, Chen SC.
Nature Communications · 2026 · two-photon lithography
Two-photon printers build optics pillar by pillar, and over a long session the pillars drift short. This one measures every pillar’s height while it cures and tops it up. Close the loop and watch the hologram sharpen.
Simulation · teaching model. Measured results further down.
The plate. 100 × 100 pixels, 7.90 µm each, as in the paper’s letter-A element. The phase design is our own Gerchberg–Saxton calculation, quantised to 11 levels (or 8 for the four-focus element). To turn phase levels into heights we assume a cured index n ≈ 1.52 at 800 nm, so the tallest A pixel is ≈ 1.40 µm. The paper gives phase levels, not these heights.
Drift. The resin’s effective sensitivity falls over the 12 simulated hours, with drift shaped like the measured open-loop trend in Fig. 4d (≈ 90% of target at 1 h, ≈ 70% at 4 h, ≈ 60% by 8 h), plus 3% pixel-to-pixel scatter. “Laser power −15%” lowers the power for two simulated hours.
Dose response. Pulses rise roughly with phase squared, as in Fig. 2d, and there is a small curing threshold (illustrative). Because of the threshold, low pillars lose proportionally more than tall ones when sensitivity drifts. The paper’s measured 1.0 µm series (Fig. 4c) also fell much faster than its 4.8 µm series.
Open loop fires the dose calibrated at hour 0. Closed loop follows the paper’s recipe: start with half the predicted dose, measure the phase (≈ 0.207 rad per µm of cured resin, with 58 nm of simulated noise), add pulses using Eq. 4 with r = 0.5, and stop within 0.015 rad. The model is updated at every new pixel, and material is never removed.
The hologram is the far field of the printed plate: the phase of each pixel in air, padded to 256 × 256 and Fourier transformed. The bright dot in the centre is the zero order, light the plate fails to steer, and it grows when heights are wrong. Pixels not yet printed carry no polymer (phase 0), so while the plate is unfinished the edge of the printed block adds a bright cross through the centre. SNR follows the paper’s definition (mean signal ÷ s.d. of background) on this simulated image, leaving out the central cross as a zero-order block would; its values are not the paper’s measurements.
Colours. Amber means within ~8% of the design height (or ~60 nm for the lowest levels). Pillars turn grey-blue as they fall short and are fully grey at 60% of the design height. Heights are drawn exaggerated so 1 µm steps are visible on 7.9 µm pixels.
Combs
Ratchet
Evidence
Instrument
Scope
Source
How the height is measured
Two combs, one cavity
A single laser cavity emits two trains of 140 fs pulses at 1052 nm. Their repetition rates (80 MHz) differ by just 356 Hz, so the pulses of one train slowly slide past the other. Each time they coincide, a short interference burst appears on the detector, about 356 times a second.
Every colour is a position
The second comb passes a grating that fans its spectrum into a line across the work field, so each colour lands at a different x. The paper calls this spectrum encoding. After the sample, a second grating recombines the colours, and the light is mixed with the first comb on a balanced detector sampled at 500 MS/s.
A Fourier transform of each burst gives the phase of every colour, and so the phase at every point along the line. Cured resin has a slightly higher refractive index (Δn ≈ 0.0347 for IP-L 780 at 1052 nm), so a structure of height h delays the light by Δφ = 2πΔn·h/λ₀ ≈ 0.207 rad per µm. Phase therefore works as a ruler for height.
Averaging longer buys precision. The paper measured a height sensitivity of 309.5 nm at 15 ms, 9.5 nm at 0.2 s and 4.8 nm at 2 s, with a lateral resolution of 1.10 µm. In-situ heights agreed with AFM after development to within 14.19 nm (mean absolute difference, 15 × 15 µm slab).
teaching animation Pulse spacing is exaggerated. The real difference is 356 Hz on 80 MHz, about 1 part in 225,000. The burst follows Eq. 1 with 64 spectral lines, and the noise levels are the paper’s measured sensitivities.
The control rule
Approach from below
Curing is irreversible: once resin is solid it cannot be un-cured. So the controller starts with half the predicted dose , measures, and then adds pulses with a first-order step, N(t+1) = N(t) + r·∇N(φt)·(φtarget − φt), with r ≤ 0.5. It stops once the measured phase is within threshold.
Why half?
The dose model is the measured link between laser pulses and phase. In Fig. 2d it is roughly quadratic, and fewer pulses are needed at higher power. If the resin has drifted to become more sensitive than the model believes, a full predicted dose overshoots, and the extra height is permanent. Starting at half the dose keeps the first exposure below the target whatever the drift. The remaining gap is then closed in shrinking steps.
After each pillar, the measured response updates the model, so the next pillar’s first guess is better. The paper updates its model at every new focus position. Its closed-loop acquisition takes 47 ms, and printing parameters are adjusted at about 1 Hz.
Try the full dose with a resin that is more sensitive than the model, or push r toward 1, and watch for a red pillar.
teaching model The red overshoot is this model’s consequence of irreversibility. The paper avoids it by starting at half the dose and setting r to 0.5 or less. Pulse counts are on the scale of Fig. 2d, the 1.0 and 4.8 µm targets are the paper’s test cuboids, and the measurement noise matches 0.2 s averaging.
Measured results
What the paper measured
Everything in this section comes from the paper’s experiments, except the last panel, which is labelled as a simulation. Grey is open loop and orange is closed loop, the paper’s own colour code.
Twelve hours of cuboids
measured Fig. 4c,d: 8 × 8 µm cuboids printed for 12 h at 45 mW, 30 mm/s and 100 nm hatching. In open loop the heights began to deviate within the first hour. Points were read from the published figure and are approximate. The mean errors are the values reported in the text.
Ten-hour cuboid array (Fig. 1c)
8.0 × 8.0 × 4.8 µm cuboids printed over 10 h. Average height error:
measured The experiments were repeated three times. Across a continuous 14-hour process making several millimetre-scale optical elements, the authors report absolute errors below 100 nm.
The letter-A hologram, printed both ways
An 11-level, 100 × 100-pixel element (7.90 µm pixels, 790 µm across) designed with the Gerchberg–Saxton algorithm to shape an 800 nm beam into a letter “A”. Both plates used the same printing parameters and post-processing. Drag across the image or use the slider to compare.
measured SNR = mean grey level ÷ s.d. of the background, as defined in the paper. Scale bars: 100 µm.
Yu et al., Nat. Commun. 17, 7135 (2026), Fig. 5d/e — CC BY 4.0; cropped, panel labels removed.
Yu et al., Nat. Commun. 17, 7135 (2026), Fig. 5g — CC BY 4.0; cropped, panel labels and scale bar removed.
A four-focus zone-plate lens
An 8-level Fresnel zone plate, 800 × 800 µm, focusing 800 nm light to four spots at f = 4 mm.
Yu et al., Nat. Commun. 17, 7135 (2026), Fig. 6a/c — CC BY 4.0; cropped, panel labels removed (focus labels F1–F4 kept in 6c).
measured Fig. 6e,f. Uniformity = (1 − (max − min)/(max + min)) × 100%.
How much error can a hologram take?
Add random, pixel-by-pixel height error to our letter-A design and look at its far field. The dial is set in phase error in the photoresist, as in the paper’s Supplementary Fig. 2, and converted to height with 0.207 rad per µm.
simulation This runs on our own design. The paper’s Supplementary Fig. 2 does the same test on its design (0–0.3 rad), and Supplementary Fig. 6 does it for the zone-plate lens. One fixed random pattern is scaled, so the image changes only when you move the dial.
The optical layout
Inside the instrument
Comb 1 is converted to 800 nm and writes. Comb 2 is spread into a rainbow line through the curing resin and is then mixed with Comb 1 to read the phase. A computer closes the loop by setting how many pulses the switch lets through. Tap or hover any part.
Redrawn and simplified from Fig. 1a, Yu et al. 2026 (CC BY 4.0). Lenses L1–L6, mirrors and fibre collimators are partly omitted.
Fabrication recipe
Clean the substrate. Glass or silicon, sonicated in acetone and then isopropanol for 10 min each, dried at 100 °C for 1 min, O2 plasma for 2 min.
Dispense resin. IP-L 780 or IP-Dip (Nanoscribe), pipetted onto the substrate. It also serves as the objective’s immersion fluid.
Write and measure. 800 nm pulses cure resin at the focus (AOM 800 kHz, galvo 2 kHz). Comb 2’s line focus reads the phase, and a precision XYZ stage (PI L-511/L-509) stitches large parts.
Develop. PGMEA for 15 min, then isopropanol for 10 min, and air-dry at room temperature.
Transparent substrates are measured in transmission and opaque ones in reflection, where the phase per µm doubles.
Yu et al., Nat. Commun. 17, 7135 (2026), Fig. 1a — CC BY 4.0; cropped.
As stated by the authors
What the loop controls, and what it assumes
The paper is specific about the job its feedback does and the conditions it relies on.
Height, not width
Only part height is feedback-controlled. Lateral size was not controlled because the 1.10 µm lateral imaging resolution is coarser than the printing resolution. The authors note that a higher-NA objective (e.g. 100×, NA 1.45) could sharpen it, at the cost of a smaller work field and shorter working distance. They also note that the ~1:2 voxel aspect ratio could be used to predict lateral size from height.
Shrinkage must be small or consistent
The loop measures parts while they are still in liquid resin. It therefore needs negligible or consistent shrinkage during development. Under their conditions, in-situ and AFM heights agreed to 14.19 nm on a 15 × 15 µm slab.
Two classes of error
The loop compensates slow drifts, such as laser power and ambient temperature, and errors that change part height, such as over- or under-exposure. The supplement lists substrate warpage, resin motion and volatility, temperature and humidity changes, and laser intensity instability.
Built on precise hardware
The method assumes repeatable, precise components: precision XYZ stages, galvo scanners and a vibration-isolation table. These are already standard in high-resolution two-photon printers. Longer averaging improves height precision at the cost of time, and faster acquisition hardware could improve both.
Source and credits
The paper
Fig. 5d and 5e (CCD images of the open- and closed-loop letter-A elements): cropped, panel labels removed, shown as a before/after wipe.
Fig. 5g (SEM close-up of the closed-loop element): cropped, labels and scale bar removed.
Fig. 6a and 6c (four foci of the open- and closed-loop zone-plate lenses): cropped, panel labels removed.
Fig. 1a (optical configuration): shown in the original and redrawn, simplified and animated as the instrument diagram.
Fig. 4c,d and 6e,f data: re-plotted from values read off the published figures (approximate) and from numbers reported in the text.
Simulations on this page. The 3D print, the hologram and its SNR, the height log, the dual-comb animation, the “approach from below” ratchet and the error dial are teaching models with illustrative parameters. Their numbers are not measurements. Measured values appear only where marked “measured”.
Typefaces: Space Grotesk and JetBrains Mono (SIL Open Font License 1.1).
Topics in this iPaper
Engineering
Subject terms: Engineering, Mechanical engineering
Abstract; Subject terms
Physics
A dual-comb heterodyne detection method is used to characterize phase changes of the structures (being printed) in photoresist
Methods, Height measurement
two photon lithography
Here we show a two-photon lithography platform powered by a single-cavity dual-comb laser that addresses this limitation through real-time correction.
Abstract
dual comb interferometry
the dual-comb-based system performs in-situ phase measurement in photoresists at 1052 nm via a spectrum-encoding technique
Introduction, final paragraph
feedback control
Figure 4a illustrates how closed-loop TPL is implemented by using the measured phases as feedback signals.
Results, Enhancement of reproducibility via closed-loop two-photon lithography
quantitative phase imaging
the measured part heights (h) have a quasi-linear relation with the phase differences
Results, Realization of in-situ nanoscale morphology measurements
diffractive optics
we designed an 11-level DOE with 100 × 100 pixels (pixel size: 7.90 µm × 7.90 µm) to shape a laser beam into a pattern of letter “A”
Results, Fabrication of large-scale optical devices via closed-loop two-photon lithography